| - | ==== Day 2026-05-31 (ln, zöe, q3k, guests) ==== | + | ==== Day 2026-06-27 (rahix) ==== |
| + | Both coaters now have their own ~3ph outlet (on the same 16A circuit). Installed secondary 4mm² PE connection for the sputter coater. | ||
| + | Checked R< | ||
| - | Second day of work on the sputter coater. | + | ==== Day 2026-06-12 (q3k, hugo, rahix) ==== |
| + | On the sputter coater, installed 3 new cooling circuit flow sensors to detect problems with the water cooling loops. | ||
| + | |||
| + | {{: | ||
| + | |||
| + | New interlock cable from those sensors to the control rack made from repurposed CAT6: | ||
| + | |||
| + | * **Orange:** Turbo Interlock (Main) | ||
| + | * **Green:** DC Interlock | ||
| + | * **Brown:** RF Interlock | ||
| + | |||
| + | Tested system, no water leaks and interlocks work. | ||
| + | |||
| + | The argon supply is still leaking close to the gas bottle. | ||
| + | |||
| + | Placed the two cheap environment sensors into the controls enclosures of the two coating systems. We did not rename them yet, so the mapping is " | ||
| + | ==== Day 2026-06-07 (q3k, guests) ==== | ||
| + | //To be documented// | ||
| + | |||
| + | ==== Day 2026-06-05 (q3k, guests) ==== | ||
| + | Hair sputtered with Cr, looks like way too much, 3mins at 400W: | ||
| + | |||
| + | {{: | ||
| + | |||
| + | Hair sputtered with Cr, 3min at 100W: | ||
| + | |||
| + | {{: | ||
| + | {{: | ||
| + | ==== Day 2026-05-2x, | ||
| + | |||
| + | Two days of work on the sputter coater. | ||
| + | |||
| + | Ar cylinder connected via needle nose valve (inlet rerouted from MFC input line into dedicated input due to lack of plumbing fittings). | ||
| RF matchbox ok, but high reflected power (4W at 100W out, 25+W at 200W out). | RF matchbox ok, but high reflected power (4W at 100W out, 25+W at 200W out). | ||
| - | Some attempts at sputtering Al with Ar work, some don't. Very inconsistent. | + | Roughing pump seems very hot, but that might be atmospheric pressure. |
| + | |||
| + | Turbo start sometimes fails and needs to be reset on the turbo controller. Roughing times out, needs set point changed. | ||
| + | |||
| + | Some attempts at sputtering Al with Ar work, some don't. Very inconsistent. Most attempts done at plasma strike pressure, ~2.2*10^-2 mbar. | ||
| + | |||
| + | Turns out RF sputter head isn't being cooled well - that might be the cause. We should dismantle it and check for damage (esp. magnets), and also try to cool it with the water chiller attached directly. | ||
| - | Turns out RF sputter | + | {{: |
| ==== Day 2026-05-16 (q3k, implr, hugo) ==== | ==== Day 2026-05-16 (q3k, implr, hugo) ==== | ||