meta data for this page
  •  

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revisionPrevious revision
Next revision
Previous revision
equipment:jeol-t330a [2024/10/16 09:45] q3kequipment:jeol-t330a [2025/05/10 22:53] (current) q3k
Line 2: Line 2:
 {{:equipment:jeol-t330a:2024-07-30-23-47-09-113.jpg?800|}} {{:equipment:jeol-t330a:2024-07-30-23-47-09-113.jpg?800|}}
  
-Our scanning electron microscope. Quick specs:+Our first scanning electron microscope. Nicknamed 'Kenji'. Quick specs:
  
 **Maximum acceleration voltage:** 30kV **Maximum acceleration voltage:** 30kV
Line 89: Line 89:
 | 3          | PK00591 EF1 pins 2, 4   | Combined blank/external video signal | | 3          | PK00591 EF1 pins 2, 4   | Combined blank/external video signal |
 | 4          | PK00492 EK3 pins 5, 7   | Video signal (post ACB)              | | 4          | PK00492 EK3 pins 5, 7   | Video signal (post ACB)              |
-| 5          |                         | GND for pin                        |+| 5          |                         | GND for pin                        |
 | 6          |                         | GND for pin 1                        | | 6          |                         | GND for pin 1                        |
 | 7          | PK00492 EK8 pin 2       | External horizontal deflection       | | 7          | PK00492 EK8 pin 2       | External horizontal deflection       |
Line 95: Line 95:
 | 9          | N/C                     | N/C                                  | | 9          | N/C                     | N/C                                  |
  
-=== Scan output connector === 
- 
-JST XH connector on a three-colored (black, green, orange) wire harness. Cobbled together by q3k, will be used for image acquisition. 
- 
-^ Pin number ^ Wire color ^ Description ^ 
-| 1          | Black      | Ground      | 
-| 2          | Green      | H or V sync | 
-| 3          | Orange     | H or V sync | 
- 
-H/V sync needs to be figured out (by following the wires or with an oscilloscope). The wires all go to U54 (opamp input) on PK00492 (the scanning board). This opamp is used to read back data from the EK9 connector. This connector allows for chaining scan/detector though the BCX x-ray unit. There might be signal integrity issues with tapping the sync data like this, we'll see. 
  
 === Connector types === === Connector types ===
Line 113: Line 103:
   - DF1 series: {{ :equipment:jeol-t330a:df1-12s-2.5r26_05_cl0541-0068-0-05_catalog_d49650_en.pdf |}}   - DF1 series: {{ :equipment:jeol-t330a:df1-12s-2.5r26_05_cl0541-0068-0-05_catalog_d49650_en.pdf |}}
  
 +==== Open Beam Interface (digital acquisition) ====
  
-===== To do =====+Use [[https://git.fa-fo.de/fafo/jeol-t330a/src/branch/main/obi/microscope.toml|microscope.toml]] from [[https://git.fa-fo.de/fafo/jeol-t330a]|our repo]]. There's also a obi.nix next to it to build the OBI codebase on Nix.
  
-==== Blockers ====+Connect the OBI to power (currently through bundle of DIN SMPSes next to OBI), connect the OBI to the WHLE-LSE-1088A box via USB (or connect directly to your PC and run your own obi-server, after editing microscope.toml). Run GUI against microscope.toml.
  
-  - Machine/acquire sample holders +TODO: move to dedicated mini PC (q3k ordered one).
-  - Track down column leak(s) +
-  - Check/replace scintillator+
  
 +===== To do =====
  
 ==== Long-term ==== ==== Long-term ====
Line 127: Line 117:
   - Replace diffusion pump with turbomolecular pump   - Replace diffusion pump with turbomolecular pump
   - Get a better roughing pump (scroll) so it doesn't spew oil everywhere   - Get a better roughing pump (scroll) so it doesn't spew oil everywhere
-  - Figure out what the 'PC' wire harness does 
-  - Digital control and acquisition 
   - E-beam lithography   - E-beam lithography
  
 +===== Log =====
  
-===== To buy =====+=== 2025/05/11 (q3k, rob, rahix) ===
  
-  - 2x M4x7.5mm screws for clamping down transformer wires in control module +Attached and tested Open Beam Interface.
-  - [[https://www.microtonano.com/Super-smooth-conductive-double-sided-adhesive-carbon-tapes-for-SEM-EDS-and-GSR.php|Double-sided carbon adhesive tape]] ~24€ +
-  - (Torx?) socket cap head M4x12 screws for vacuum fittings+
  
-===== Log =====+=== 2025/01/31 (q3k, rob, 0ff) === 
 + 
 +New sample holder (EM-Tec JV50 + EM-Tec J4) has been attached. 
 + 
 +=== 2025/01/10 (q3k, rob, 0ff) === 
 + 
 +New scintillator (PLANOTEC P47) has been mounted. 
 + 
 +=== 2024/11/09 (hugo, rahix, leah, ln) === 
 + 
 +Replaced the succbone relay boards with a MODBUS-based system.  Added MODBUS-based temperature measurements for the diffusion pump.
  
 === 2024/10/15 (q3k, rob) === === 2024/10/15 (q3k, rob) ===