Line 2: | Line 2: | ||
{{: | {{: | ||
- | Our scanning electron microscope. Quick specs: | + | Our first scanning electron microscope. Nicknamed ' |
**Maximum acceleration voltage:** 30kV | **Maximum acceleration voltage:** 30kV | ||
Line 89: | Line 89: | ||
| 3 | PK00591 EF1 pins 2, 4 | Combined blank/ | | 3 | PK00591 EF1 pins 2, 4 | Combined blank/ | ||
| 4 | PK00492 EK3 pins 5, 7 | Video signal (post ACB) | | | 4 | PK00492 EK3 pins 5, 7 | Video signal (post ACB) | | ||
- | | 5 | | GND for pin 6 | | + | | 5 | | GND for pin 4 | |
| 6 | | GND for pin 1 | | | 6 | | GND for pin 1 | | ||
| 7 | PK00492 EK8 pin 2 | External horizontal deflection | | 7 | PK00492 EK8 pin 2 | External horizontal deflection | ||
Line 104: | Line 104: | ||
| 3 | Orange | | 3 | Orange | ||
- | H/V sync needs to be figured out (by following the wires or with an oscilloscope). The wires all go to U54 (opamp input) on PK00492 (the scanning board). This opamp is used to read back data from the EK9 connector. This connector allows for chaining scan/ | + | H/V sync needs to be figured out (by following the wires or with an oscilloscope). |
+ | |||
+ | H/V sync values are ramps +/- 5V (opamps are supplied +/- 8V). | ||
+ | |||
+ | The wires all go to U54 (opamp input) on PK00492 (the scanning board). This opamp is used to read back data from the EK9 connector. This connector allows for chaining scan/ | ||
=== Connector types === | === Connector types === | ||
Line 115: | Line 119: | ||
===== To do ===== | ===== To do ===== | ||
- | |||
- | ==== Blockers ==== | ||
- | |||
- | - Machine/ | ||
- | - Track down column leak(s) | ||
- | - Check/ | ||
- | |||
==== Long-term ==== | ==== Long-term ==== | ||
Line 127: | Line 124: | ||
- Replace diffusion pump with turbomolecular pump | - Replace diffusion pump with turbomolecular pump | ||
- Get a better roughing pump (scroll) so it doesn' | - Get a better roughing pump (scroll) so it doesn' | ||
- | - Figure out what the ' | ||
- Digital control and acquisition | - Digital control and acquisition | ||
- E-beam lithography | - E-beam lithography | ||
Line 135: | Line 131: | ||
- 2x M4x7.5mm screws for clamping down transformer wires in control module | - 2x M4x7.5mm screws for clamping down transformer wires in control module | ||
- | - [[https:// | ||
- (Torx?) socket cap head M4x12 screws for vacuum fittings | - (Torx?) socket cap head M4x12 screws for vacuum fittings | ||
===== Log ===== | ===== Log ===== | ||
+ | |||
+ | === 2025/01/31 (q3k, rob, 0ff) === | ||
+ | |||
+ | New sample holder (EM-Tec JV50 + EM-Tec J4) has been attached. | ||
+ | |||
+ | === 2025/01/10 (q3k, rob, 0ff) === | ||
+ | |||
+ | New scintillator (PLANOTEC P47) has been mounted. | ||
+ | |||
+ | === 2024/11/09 (hugo, rahix, leah, ln) === | ||
+ | |||
+ | Replaced the succbone relay boards with a MODBUS-based system. | ||
=== 2024/10/15 (q3k, rob) === | === 2024/10/15 (q3k, rob) === |