equipment:zeiss-dsm-962 [2025/09/20 14:56] – q3k | equipment:zeiss-dsm-962 [2025/09/28 17:15] (current) – Describe vibration isolation rahix | ||
---|---|---|---|
Line 28: | Line 28: | ||
- The system should now be operational. | - The system should now be operational. | ||
- | ==== Venting the column === | + | ==== Venting the column |
To vent the column (eg. to change samples or inspect/ | To vent the column (eg. to change samples or inspect/ | ||
Line 36: | Line 36: | ||
- Open chamber / gun, do your work, close it. | - Open chamber / gun, do your work, close it. | ||
- Press the EVAC button on the vacuum control panel. | - Press the EVAC button on the vacuum control panel. | ||
+ | |||
+ | ==== Setting up the vibration isolation ==== | ||
+ | |||
+ | Before imaging, you will need to set up the vibration isolation of the SEM column. | ||
+ | |||
+ | - At the back of the SEM you will find 3 translucent blue tubes, each ending in a tire valve adapter. | ||
+ | {{: | ||
+ | - Connect the bike pump to the tube that runs deepest into the SEM, connecting to the front isolation bellows (this tube is split to connect to both front bellows at the same time). | ||
+ | - Pump up the bellows to 2.9 barg. After reaching that pressure, the bellows start suddenly moving quickly. | ||
+ | - Immediately stop pumping and go very slowly. | ||
+ | - After pumping up the front bellows, attach the bike pump to the back left bellow. | ||
+ | - Repeat the pumping procedure detailed above. | ||
+ | - Finally pump up the back right bellow using the same method. | ||
+ | - Over time, some of the tire valves may leak some air. Just refill the bellows as needed. | ||
+ | |||
+ | You can find more info in the [[equipment: | ||