 ===== Journal for the Edwards AUTO 306 Coating Systems ===== ===== Journal for the Edwards AUTO 306 Coating Systems =====
 This is the journal.  See the [[equipment:edwards-306:start|Edwards AUTO 306 Coating System]] project page for an overview. This is the journal.  See the [[equipment:edwards-306:start|Edwards AUTO 306 Coating System]] project page for an overview.
 +
 +==== Day 2026-06-27 (rahix) ====
 +Both coaters now have their own ~3ph outlet (on the same 16A circuit). Installed secondary 4mm² PE connection for the sputter coater.
 +Checked R<sub>LOW</sub> and R<sub>ISO</sub> for both coaters.
  
 ==== Day 2026-06-12 (q3k, hugo, rahix) ==== ==== Day 2026-06-12 (q3k, hugo, rahix) ====
Line 13: Line 17:
   * **Brown:** RF Interlock   * **Brown:** RF Interlock
  
-Tested system, no leaks and interlocks work.+Tested system, no water leaks and interlocks work
 + 
 +The argon supply is still leaking close to the gas bottle.  Will need to fix that soon. 
 + 
 +Placed the two cheap environment sensors into the controls enclosures of the two coating systems. We did not rename them yet, so the mapping is "Unusually Large Staircase" -> Sputter Coater and "Chillzone" -> Evap Coater.
 ==== Day 2026-06-07 (q3k, guests) ==== ==== Day 2026-06-07 (q3k, guests) ====
 //To be documented// //To be documented//