| > //This page is a lab notebook. It documents the work done in the lab and our thoughts, theories, and findings related to it. Everything you read here is work in progress and not final results. We believe it is best to have our lab notes out in the open, because it gives more detailed insight into what worked for us and what didn' | > //This page is a lab notebook. It documents the work done in the lab and our thoughts, theories, and findings related to it. Everything you read here is work in progress and not final results. We believe it is best to have our lab notes out in the open, because it gives more detailed insight into what worked for us and what didn' | ||
| + | ==== Day 2026-08-14 (rahix, q3k, hugo) ==== | ||
| + | Used the sputter coater again today to coat another hair with chromium. | ||
| + | |||
| + | During pump down, we ran into the penning gauge '' | ||
| + | |||
| + | We didn't check and decided to go ahead with coating regardless, flying blind on the pressure inside the chamber. | ||
| + | |||
| + | Coating went well, the hair turned out great at 30s coating time. | ||
| + | |||
| + | Will need to check the penning gauge //again// next time. At final pump down, the vacuum system repeatedly failed with '' | ||
| + | |||
| + | ==== Day 2026-08-07 (rahix, guest) ==== | ||
| + | Measured the sample mounting rails inside both coaters to design a custom fixturing system to hold samples above the coating heads. | ||
| + | |||
| + | ==== Day 2026-08-01 ==== | ||
| + | A lot of successful sputter coating during the third //Find-Out Forum//. | ||
| + | |||
| + | ==== Day 2026-07-25 (q3k) ==== | ||
| + | Today, the sputter coater failed due to penning gauge error: | ||
| + | |||
| + | {{: | ||
| + | |||
| + | Power cycle did not help so decided to take apart the system. Penning gauge sits all the way in the back (the red device sticking down from the vacuum chamber). Suspicion was confirmed: There was debris inside, likely causing it to fault. Opened the gauge and cleaned everything. | ||
| + | |||
| + | {{: | ||
| + | {{: | ||
| + | |||
| + | After reassembly, it works again! Measuring 2.8e-4 mbar. | ||
| + | |||
| + | ----- | ||
| + | |||
| + | Worked on building automated control of the MFC (Mass Flow Controller). It is now connected to a MODBUS Analog Output module. This can then be controlled from the lab network via a MODBUS-TCP bridge. | ||
| + | |||
| + | TODO q3k: Link script | ||
| + | |||
| + | This is still work in progress and will likely get replaced by a proper control system (soon< | ||
| ==== Day 2026-07-17 (q3k, guests) ==== | ==== Day 2026-07-17 (q3k, guests) ==== | ||