| ===== Journal for E306A Coating System ===== | ===== Journal for E306A Coating System ===== | ||
| - | ==== Day 2026-03-30 ==== | + | See [[equipment: |
| - | Brought | + | |
| + | ==== Day 2026-04-25 (rahix) | ||
| + | Electrical inspection (V3) of the machine. Rewired a few plugs. Everything passed! | ||
| + | ==== Day 2026-04-12 (q3k) ==== | ||
| + | |||
| + | Fit new vent valve holder. Still a bit loose but better. Should probably hold the pipeline instead of the valve... | ||
| + | |||
| + | Reprinted knob to fit shaft better. Partially retapped shaft to M6 - previous thread didn't seem metric. | ||
| + | |||
| + | Fixed leak at pirani gauge after swapping it for some tests earlier. Took normal KF centering ring. | ||
| + | |||
| + | Attempted to run diffusion pump. A bunch of stuff offgassed and the pump sizzled but there was no extra vacuum. Need to inspect and probably fill, no other choice now... | ||
| + | |||
| + | ==== Day 2026-04-04 (q3k, Hugo, Rob) ==== | ||
| + | |||
| + | Ordered activated alumina from Roth. | ||
| + | |||
| + | Valve actuates, but leaks. Ended up replacing it with a modern vent valve with an ISO KF flange. The threaded flanges from Edwards are compatible enough to get us to a decent rough vacuum. New valve needs wiring done well (currently there is no strain relief) and a new mount (currently everything dangles off the vent line, which risks breaking it). | ||
| + | |||
| + | Water inlet block got PTFE tape replaced with o-rings. Water outlet tee got swapped for one borrowed from the membrane pump. | ||
| + | |||
| + | Roughing pump oil is really smelly, even with the oil mist filter. We really should replace it. | ||
| + | |||
| + | Replaced oil with ISO VG 68 from 2022. Did a small purge cycle just to get the manky old oil out. Now reaching an ultimate pressure of around 8e-2 mbar. Put old oil into plastic canister, ended up accidentally causing an emulsion with water using detergent - we should clean that up (salt it out with brine?) before handing it off to the Wertstoffhof. | ||
| + | |||
| + | 3D printed a holder for the new vent valve, but needs some small rework to hold the valve better. | ||
| + | |||
| + | ==== Day 2026-04-03 (q3k, Rob, Hugo) ==== | ||
| + | |||
| + | Got system down to a rough vacuum (1e-1 mbar). | ||
| + | |||
| + | Fitted custom 3d printed knob/ | ||
| + | |||
| + | Discovered how the valve blocks work. Mostly understand the vacuum system. Based around mostly ISO KF compatible flanges but with proprietary screw-based ' | ||
| + | |||
| + | Found film thickness monitor oscillator and quartz balance. Quartz crystal parameters are documented, we should be able to use an arbitrary frequency counter as a film thickness monitor. | ||
| + | |||
| + | Discovered air vent valve is stuck closed. Giving it 230V directly does not activate it. Took apart, coils measure fine, rebuilt mechanicals (screws were rusted but Al block seems fine), piston moves. Started REing electronic control system to be able to diagnose and fix it. Limit switches are unused by the control system. | ||
| + | |||
| + | Discovered foreline trap is empty - needs 2mm activated alumina dessicant charge. Started looking for suppliers. | ||
| + | |||
| + | After manually venting (by removing parts of the vacuum system) was able to lift bell jar. Cleaned seal with IPA and kimwipes. | ||
| + | |||
| + | Started inspecting vacuum chamber fixtures. | ||
| + | |||
| + | Substrate/ | ||
| ==== Day 2026-04-02 (q3k, Rob, Hugo) ==== | ==== Day 2026-04-02 (q3k, Rob, Hugo) ==== | ||
| Line 12: | Line 57: | ||
| Rebuilt inlet block, replaced push fitting in it. Doesn' | Rebuilt inlet block, replaced push fitting in it. Doesn' | ||
| - | Bypassed inlet coupler and outlet coupler (connecting water cooling circuit directly to block and output tee). | + | Bypassed inlet coupler and outlet coupler (connecting water cooling circuit directly to block and output tee). Still leaks at tee, ordered replacement. |
| + | |||
| + | Found dangling 230V cable (black) in lower chassis. Discovered it's a feedback cable (230V, NO, relay input) to the electric shutter from a missing film thickness monitor. Took apart electric shutter controller to discover this, pulled N wire from its terminal block by accident, reinstalled. Should be fine, but would really benefit from a re-crimp with ferrules. | ||
| - | Found dangling 230V cable (black) in lower chassis. Discovered it's a feedback cable to the electric shutter from a missing film thickness monitor. Took apart electric shutter controller to discover this, pulled N wire from terminal block by accident, reinstalled. Should be fine. | + | Briefly |
| - | + | ||
| - | Quickly | + | |
| Installed 3d printed toilet paper roll oil mist filter. Might prevent the front panel from closing. | Installed 3d printed toilet paper roll oil mist filter. Might prevent the front panel from closing. | ||
| + | |||
| + | ==== Day 2026-03-30 ==== | ||
| + | Brought the system into out lab. | ||
| + | |||
| + | Bypassed front panel lock (no keys), replaced key lock core with flathead-driven lock core. | ||