meta data for this page
  •  

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revisionPrevious revision
Next revision
Previous revision
equipment:zeiss-dsm-962:log [2026/02/14 17:34] โ€“ [Lab Log for the Zeiss DSM "Kevin"] rahixequipment:zeiss-dsm-962:log [2026/03/01 19:21] (current) โ€“ [Lab Log for the Zeiss DSM "Kevin"] rahix
Line 1: Line 1:
 ====== Lab Log for the Zeiss DSM "Kevin" ====== ====== Lab Log for the Zeiss DSM "Kevin" ======
  
 +==== Day 2026-03-01 (rahix) ====
 +Tried connecting ''FRONT'', ''BACK1'', and ''BACK2'' all together using pneumatic T-couplers.  It doesn't work, the weight of the column is too much off the side, meaning one bellow stays in contact with the bottom while the diagonally opposite one is at the top limit.
 +
 +Reverted back to the previous state with three separate valves.
 ==== Day 2026-02-14 (rahix) ==== ==== Day 2026-02-14 (rahix) ====
 Fixed the very leaky tire valves of the isolation system. We now have proper pneumatic isolation valves on all three tubes instead.  They are labelled now: Fixed the very leaky tire valves of the isolation system. We now have proper pneumatic isolation valves on all three tubes instead.  They are labelled now:
Line 9: Line 13:
  
 {{:equipment:zeiss-dsm-962:20260214_labelled-bellows.png?400|}} {{:equipment:zeiss-dsm-962:20260214_labelled-bellows.png?400|}}
 +
 +Updated the instructions here: [[equipment:zeiss-dsm-962#setting_up_the_vibration_isolation]]
 ==== Day 2026-01-24 (q3k, rahix, rob, hugo) ==== ==== Day 2026-01-24 (q3k, rahix, rob, hugo) ====