 ====== Lab Log for the Zeiss DSM "Kevin" ====== ====== Lab Log for the Zeiss DSM "Kevin" ======
  
 +==== Day 2026-03-01 (rahix) ====
 +Tried connecting ''FRONT'', ''BACK1'', and ''BACK2'' all together using pneumatic T-couplers.  It doesn't work, the weight of the column is too much off the side, meaning one bellow stays in contact with the bottom while the diagonally opposite one is at the top limit.
 +
 +Reverted back to the previous state with three separate valves.
 ==== Day 2026-02-14 (rahix) ==== ==== Day 2026-02-14 (rahix) ====
 Fixed the very leaky tire valves of the isolation system. We now have proper pneumatic isolation valves on all three tubes instead.  They are labelled now: Fixed the very leaky tire valves of the isolation system. We now have proper pneumatic isolation valves on all three tubes instead.  They are labelled now: