+ | ⇒ [[equipment: | ||
==== Powering on the scope ==== | ==== Powering on the scope ==== | ||
Line 27: | Line 28: | ||
- The system should now be operational. | - The system should now be operational. | ||
- | ==== Powering off the system | + | ==== Venting the column === |
+ | |||
+ | To vent the column (eg. to change samples or inspect/ | ||
+ | |||
+ | - Press the VENT button on the vacuum control panel | ||
+ | - Wait a surprisingly long time (until hissing is done) | ||
+ | - Open chamber / gun, do your work, close it. | ||
+ | - Press the EVAC button on the vacuum control panel. | ||
+ | |||
+ | ==== Using the EDX system ==== | ||
+ | |||
+ | The detector is currently not mounted, but you can try using the system for image acquisition. | ||
+ | |||
+ | - Power on the Dell PC, wait for a boot to desktop | ||
+ | - Start the 'NSS ohne drift' script from the desktop | ||
+ | - Start the ' | ||
+ | - Select an existing or create a new project | ||
+ | - Ignore the warning about the detector being at high temperature (it's not, it's just disconnected) | ||
+ | - Select ' | ||
+ | - Press play to do a live acquisition, | ||
+ | |||
+ | If everything goes well, the DSM will go into ' | ||
+ | |||
+ | ==== Powering off the scope ==== | ||
- Turn the key to the ' | - Turn the key to the ' |