meta data for this page
Differences
This shows you the differences between two versions of the page.
| Both sides previous revisionPrevious revisionNext revision | Previous revision | ||
| equipment:zeiss-dsm-962 [2025/09/20 14:56] โ q3k | equipment:zeiss-dsm-962 [2026/02/14 17:37] (current) โ [Setting up the vibration isolation] rahix | ||
|---|---|---|---|
| Line 28: | Line 28: | ||
| - The system should now be operational. | - The system should now be operational. | ||
| - | ==== Venting the column === | + | ==== Venting the column |
| To vent the column (eg. to change samples or inspect/ | To vent the column (eg. to change samples or inspect/ | ||
| Line 36: | Line 36: | ||
| - Open chamber / gun, do your work, close it. | - Open chamber / gun, do your work, close it. | ||
| - Press the EVAC button on the vacuum control panel. | - Press the EVAC button on the vacuum control panel. | ||
| + | |||
| + | ==== Setting up the vibration isolation ==== | ||
| + | |||
| + | Before imaging, you will need to set up the vibration isolation of the SEM column. | ||
| + | |||
| + | - At the back of the SEM you will find 3 translucent blue tubes, each ending in an isolation valve and a tire valve adapter. | ||
| + | {{: | ||
| + | - Connect the bike pump to the tube labelled '' | ||
| + | - Pump up the bellows to 2.9 barg. After reaching that pressure, the bellows start suddenly moving quickly. | ||
| + | - Immediately stop pumping and go very slowly. | ||
| + | - Close the blue isolation valve on the tube and disconnect the bike pump. Carefully release some air using the isolation valve until the bellows height is about half-way. | ||
| + | - You want the bellows to be roughly around the center of their extension range. | ||
| + | - After pumping up the front bellows, attach the bike pump to one of the back bellows. | ||
| + | - Repeat the pumping procedure detailed above. | ||
| + | - Finally pump up the other back bellow ('' | ||
| + | |||
| + | You can find more info in the log entries ([[equipment: | ||