meta data for this page
Differences
This shows you the differences between two versions of the page.
Both sides previous revisionPrevious revisionNext revision | Previous revision | ||
projects:stm:log [2025/09/07 22:49] – [Day 2025-09-07 (rahix)] rahix | projects:stm:log [2025/09/30 22:27] (current) – Remove nolink rahix | ||
---|---|---|---|
Line 1: | Line 1: | ||
====== Lab Notes for the Scanning Tunneling Microscope ====== | ====== Lab Notes for the Scanning Tunneling Microscope ====== | ||
+ | ==== Day 2025-09-20 (rahix) ==== | ||
+ | Modelled more of the as-built of the current STM setup in mechanical CAD so we have a reference model: | ||
+ | |||
+ | {{: | ||
==== Day 2025-09-07 (rahix) ==== | ==== Day 2025-09-07 (rahix) ==== | ||
Took the video capture from yesterday, of the beam bouncing from the STM off a mirror 2.35m away and back to the camera on the STM and ran some analysis on the data. | Took the video capture from yesterday, of the beam bouncing from the STM off a mirror 2.35m away and back to the camera on the STM and ran some analysis on the data. | ||
- | Analysis works like this: For each frame, we find the beam center. | + | Analysis works like this: For each frame, we find the beam center. |
Here is the final figure: | Here is the final figure: | ||
Line 513: | Line 517: | ||
Played a bit more with the [[https:// | Played a bit more with the [[https:// | ||
- | {{: | + | {{: |
Line 530: | Line 534: | ||
Calculated the damping behavior of our setup: | Calculated the damping behavior of our setup: | ||
- | {{: | + | {{: |
EMI Box: | EMI Box: | ||
Line 607: | Line 611: | ||
Removed the formwork from the conrete block we poured on friday. | Removed the formwork from the conrete block we poured on friday. | ||
- | {{: | + | {{: |
* Lessons learned: | * Lessons learned: | ||
Line 625: | Line 629: | ||
Poured a concrete weight for the vibration isolation system: | Poured a concrete weight for the vibration isolation system: | ||
- | {{: | + | {{: |
* Lessons learned: | * Lessons learned: |