equipment:edwards-306:log [2026/04/19 00:55] – [Day 2026-04-18 (q3k, Hugo, Rob, Jana, Guest x3)] q3kequipment:edwards-306:log [2026/04/19 11:55] (current) q3k
Line 16: Line 16:
 Interlock form control box didn't seem to enable turbo controller (lines 13/14 on I/O connector). Bypassed for now, cable marked. Interlock form control box didn't seem to enable turbo controller (lines 13/14 on I/O connector). Bypassed for now, cable marked.
  
-Main valve didn't seem to be actuated from the control box. Manually enabled for now.+Vent valve didn't seem to be actuated from the control box. Manually enabled for now.
  
 System went down to vacuum but not without a fight (timeouts, etc. - needs a good pumpdown for a few hours). Cleaned main seal. Managed to strike a plasma via HT. Managed to blow out thermal evap boat via LT. System went down to vacuum but not without a fight (timeouts, etc. - needs a good pumpdown for a few hours). Cleaned main seal. Managed to strike a plasma via HT. Managed to blow out thermal evap boat via LT.