| > //This page is a lab notebook. It documents the work done in the lab and our thoughts, theories, and findings related to it. Everything you read here is work in progress and not final results. We believe it is best to have our lab notes out in the open, because it gives more detailed insight into what worked for us and what didn' | > //This page is a lab notebook. It documents the work done in the lab and our thoughts, theories, and findings related to it. Everything you read here is work in progress and not final results. We believe it is best to have our lab notes out in the open, because it gives more detailed insight into what worked for us and what didn' | ||
| + | |||
| + | ==== Day 2026-07-18 (q3k, rahix) ==== | ||
| + | For reference, this is the current sample configuration in the SEM: | ||
| + | |||
| + | {{: | ||
| + | |||
| + | ==== Day 2026-07-16 (q3k, guest) ==== | ||
| + | Finally replaced the o-ring. In the process, the interlock switch for the top of the column broke. It is jumpered out for now. | ||
| + | |||
| + | * Original switch: [[https:// | ||
| + | * Possible replacement: | ||
| + | |||
| + | ==== Day 2026-07-11 (q3k, rahix) ==== | ||
| + | Started work on designing a proper enclosure for the OBI. | ||
| + | ==== Day 2026-07-10 (q3k) ==== | ||
| + | Continued work on the OBI. Here's a picture acquired through it: | ||
| + | |||
| + | {{: | ||
| + | |||
| + | ==== Day 2026-07-08 (q3k, hugo) ==== | ||
| + | New tool worked to get the locknut out. O-ring indeed got destroyed; ordered replacements (it's a 6×2mm). In the meantime, sealed it up again and while not pumping down great, we can still get a good enough vacuum for imaging. | ||
| + | |||
| + | Installed the beam blanker: | ||
| + | |||
| + | {{: | ||
| + | |||
| + | It has an SMC connector, will need to get an appropriate cable. | ||
| ==== Day 2026-07-05 (rahix, rob) ==== | ==== Day 2026-07-05 (rahix, rob) ==== | ||