 Removed the beam-blanker again since we suspect it creates a weirdly shaped beam image (the emission image looks 0ff ever since we installed it). Removed the beam-blanker again since we suspect it creates a weirdly shaped beam image (the emission image looks 0ff ever since we installed it).
  
-Disassemled the inner liner tube again to check it for suspected missing apertures.+Disassembled the inner liner tube again to check it for suspected missing apertures - we are indeed missing 2x 3mm (OD) apertures as well as 2x retaining clips.
  
 +Checked for the weird artifacts and couldn't see them as strong as before, but they are not gone. Emission image looks like a half moon shape still.
 ==== Day 2026-07-18 (q3k, rahix) ==== ==== Day 2026-07-18 (q3k, rahix) ====
 For reference, this is the current sample configuration in the SEM: For reference, this is the current sample configuration in the SEM: